Decision rule · guard band Decision rule calculator: conforming, undecidable or out?

A reading of 25.018 against a limit of ±0.02 looks fine – until the certificate says U = 0.004. Enter nominal, reading, limit and U; the verdict and the arithmetic behind it update as you type. No sign-up, nothing leaves your browser.

Last updated: November 6, 2026 · Reviewed by irot

  • Free for up to 20 devices
  • No payment details
  • Interface in English and German
Uncertainty
Measurement uncertainty in Kalibrat

The calculator

All values in the same unit (for example mm). The limit is the permitted deviation from the nominal value, as a positive number.

How the verdict is reached

Conforming
|deviation| + U is within the limit. The true deviation, wherever it lies within the uncertainty, is inside.
Close to the limit
Conforming, but |deviation| + U reaches 80 % of the limit. Worth a note – and a shorter interval if it happens repeatedly.
Undecidable
The reading is inside, but |deviation| + U is above the limit: the true deviation could be on either side. Conformity is not demonstrated. Measure with less uncertainty, or decide according to your own documented rule.
Non-conforming
|deviation| − U is above the limit: outside even in the best case.

This is the two-zone approach of ISO 14253-1, in our own words. With simple acceptance the reading alone decides – which is only reasonable while the uncertainty is small against the limit; the calculator applies the common 4 : 1 ratio and falls back to the rule with uncertainty when U is larger than a quarter of the limit. Without a U, the reading alone decides under either rule. Neither rule accepts anything the plain comparison would reject.

Same verdict as in the Kalibrat register, where the rule is a setting and is stored with each calibration record, so that an old result is not re-judged when the setting changes later.

FAQ

Frequently asked questions

What is a decision rule?

The written rule for how measurement uncertainty is taken into account when a result is compared with a limit. Accredited laboratories state the rule on the certificate; your company should state its own in the quality system.

What is a guard band?

The strip inside the limit by which the acceptance zone is narrowed – here equal to U. A reading must stay inside the limit by at least U to count as conforming.

Where does the 4 : 1 ratio come from?

A long-standing rule of thumb: if the uncertainty is at most a quarter of the tolerance, ignoring it changes very few verdicts. Above that, it must enter the decision.

Is my data uploaded?

No. The calculation runs entirely in your browser.